Electrothermic factors optimization in electrothermal atomic absorption spectrometry via an optimal experimental design matrix - Université de Bretagne Occidentale
Journal Articles Analytica Chimica Acta Year : 1998

Electrothermic factors optimization in electrothermal atomic absorption spectrometry via an optimal experimental design matrix

Abstract

A new approach is proposed to optimize some instrumental electrothermic parameters, i.e. injected volume, calcination duration, calcination and atomization temperatures, gas flow, in electrothermal atomic absorption spectrometry. It is based on the monitoring of the detection limit estimation variation by using an unconventional experimental design well adapted to a particular experimental domain. This method was tested on Cd, Mn and Cu in sea water. By comparison with the default operating conditions, the limits of detection were improved in some cases and always obtained on carrying out less experiments than usual. Copyright (C) 1998 Elsevier Science B.V.

Dates and versions

hal-01770753 , version 1 (19-04-2018)

Identifiers

Cite

H. Le Garrec, P. Giamarchi, Jean Yves Cabon, A. Le Bihan. Electrothermic factors optimization in electrothermal atomic absorption spectrometry via an optimal experimental design matrix. Analytica Chimica Acta, 1998, 368, pp.59--70. ⟨10.1016/S0003-2670(98)00178-0⟩. ⟨hal-01770753⟩